SLEEP
SLEEP electrode cap is the textile cap with pre-installed Ag/AgCl sintered MCScap‑T electrodes and with DB‑25M common connector. The electrode cap is designed for non-invasive registration EEG and use of electroencephalographs and other amplifiers biological signals. The electrode cap is a reusable medical device.
Textile cap is made of elastic material, preserving the shape and size. The cap provides the exact position of the electrodes on the head without additional measurements and adjustments. Large holes are provided for ventilation and access to the electrodes and patient’s skin. The cap is fixed on the head with the chin or chest belt. The caps are marked according to the 10-10 system. Size identification is carried out by the color of the material or by the color of the seam.
MCScap‑T is a thin cup Ag / AgCl sintered electrode for EEG recording. MCScap‑T is designed for maximum patient comfort during examination in lying down position, for example, during sleep EEG or for examination of newborns. The design of this electrode is the most preferred for conducting combined TMS-EEG studies. The Ag / AgCl sintered electrode material guarantees minimum polarization and long-term signal stability, as well as an increased electrode life. The conductive surface of the MCScap electrodes is not in direct contact with the skin. Contact is provided by a conductive substance. A hole in the electrodes is provided to add a conductive gel. Use with conductive paste is possible. Electrode have additional labeling what makes easy to rearrange them to new textile base.
Electrode | MCScap‑T |
Material of electrode conductive surface | Ag/AgCl sintered |
Electrode body material | polyurethane |
Square of electrode conductive surface | 7 mm2 |
Internal diameter of the electrode at the point of contact of the electrode contact substance with the skin | 6 mm |
Surface area of contact of the electrode substance with the skin | 28.3 mm2 |
Outer diameter of the electrode at the point of contact with the skin | 11.5 mm |
Distance from the skin to the electrode conductive surface | 1.6 mm |
The diameter of the hole in the electrode to add gel | 2.8 mm |
Electrode polarization | ≤ 50 mV |
Resistance of electrodes insulation | ≥ 1000 MΩ |
Dielectric strength of electrodes insulation | 1500 V |
The impedance of the electrode | ≤ 5 kΩ |
Number of electrodes / channels | 21 / 20 |
Electrode positions | Fp1, Fp2, F3, F4, C3, C4, P3, P4, O1, O2, F7, F8, T7(Т3), T8(T4), P7(T5), P8(T6), Cz, Fz, Pz, GND, Ref |
Connector type | DB‑25M |
Cable length | 1.5 m |
Location of the output of the electrode cable from the cap | top of the head |
Ear electrodes | no |
Marking of the textile cap | yes |
Marking of the electrodes | yes |
Weight of EEG cap | < 250 g |
Net weight | < 750 g |
-
SLEEP electrode cap,
-
EEG starter kit
-
User Manual.
Size | Application | Head circumference | Textile or seam color |
---|---|---|---|
XL | Adult | 60-66 cm | Green |
XL/L | Adult | 57-63 cm | Green + blue |
L | Adults (most) | 54-60 cm | Blue |
L/M | Teens, adults | 51-57 cm | Blue + red |
M | Children, teens | 48-54 cm | Red |
M/S | Children under 5 years old | 45-51 cm | Red + yellow |
S | Children under 2 years old | 42-48 cm | Yellow |
S/XS | Children under 1 year old | 39-45 cm | Yellow + green |
XS | Children up to 4 months | 36-42 cm | Green |
Inf I | Newborns before the 1st month | 32-36 cm | Blue |
Inf II | Newborns, premature | 28-32 cm | Red |
Inf III | Newborns, premature | 24-28 cm | Yellow |